Título : Avaliação do desempenho da tecnologia NeoSoniX TM em cirurgias de catarata
Otros títulos : Performance evaluation of NeoSoniX TM technology in cataract surgery
Autor : Akaishi, Leonardo
Silva, Rodrigo Vaz de Queiroz
Assunto:: Catarata
Facoemulsificação
Instrumentos e aparelhos cirúrgicos
Estudo comparativo
Fecha de publicación : 2006
Editorial : Conselho Brasileiro de Oftalmologia
Citación : AKAISHI, Leonardo; SILVA, Rodrigo Vaz de Queiroz. Avaliação do desempenho da tecnologia NeoSoniX TM em cirurgias de catarata. Arquivos Brasileiros de Oftalmologia, v. 69, n. 3, p. 389-393, 2006. DOI: https://doi.org/10.1590/S0004-27492006000300019. Disponível em: https://www.scielo.br/j/abo/a/bSL8LF9FSdvMS3bDX8qPV9B/?lang=pt#. Acesso em: 10 ago. 2021.
Resumen : OBJETIVOS: As técnicas e resultados da cirurgia de catarata melhoraram drasticamente com o avanço tecnológico. Um dos motivos para melhoria seria a menor utilização de ultra-som com novos aparelhos de facoemulsificação. O objetivo deste estudo é confirmar na prática a vantagem teórica da caneta NeoSoniX TM. MÉTODOS: Foi realizada uma análise na qual os autores compararam resultados dos parâmetros registrados pelo aparelho de facoemulsificação Legacy durante 300 cirurgias de catarata realizadas com caneta de ultra-som (handpiece) convencional e 100 com a caneta NeoSoniX TM. Todas as cirurgias foram realizadas pelo mesmo cirurgião, com a mesma técnica cirúrgica. A escolha da "handpiece" foi aleatória, sendo comparadas, em relação ao grau de dureza das cataratas, os parâmetros de tempo do ultra-som, quantidade de solução salina balanceada e porcentagem de "average power" utilizadas durante as cirurgias. RESULTADOS: Foi realizada análise retrospectiva numa série cronológica de casos, mostrando que em qualquer que seja o grau de dureza da catarata, o tempo de utilização do ultra-som com a caneta NeoSoniX foi inferior ao da caneta convencional (p<0,000001). O volume de solução salina balanceada utilizado durante as cirurgias foi menor a favor da tecnologia NeoSoniX (p<0,000001). Em relação à porcentagem de potência do ultra-som utilizada com as diferentes tecnologias, encontrou-se valor inferior no grupo da caneta NeoSoniX (p<0,000001). CONCLUSÕES: Houve superioridade significativa no uso da caneta NeoSoniX comparada à caneta convencional do aparelho Legacy nas cirurgias de catarata, independente do seu grau.
Abstract: PURPOSE: The techniques and results of cataract surgery have improved dramatically with technological advancements. One of the reasons for this improvement would be the reduced use of ultrasound with the new phacoemulsification devices. The objective of this study is to confirm the theoretical advantage of the NeoSoniX handpiece. METHODS: The authors compared the results of the parameters recorded by the Legacy phacoemulsification system during 300 cataract surgeries performed with the use of the conventional ultrasound handpiece and 100 performed with the innovative NeosonixTM. All surgeries were performed by the same senior surgeon, using the same surgical technique. The choice of handpiece was random. The handpieces were compared considering cataract grade of hardness, in relation to parameters of ultrasound time, amount of balanced saline solution, and percentage of average power used during surgery. RESULTS: The analysis performed retrospectively on a chronological series of cases showed that the time of ultrasound used with the NeoSoniX handpiece was inferior to that with the conventional handpiece, for any cataract grade of hardness (p<0.000001). The volume of balanced saline solution used during surgical procedures was smaller, favoring the NeoSoniX technology (p<0.000001). In relation to the percentage of ultrasound energy used with the different technologies, an inferior value was found in the group who underwent surgery with the use of the NeoSoniX handpiece (p<0.000001). CONCLUSIONS: Based on the information provided by these surgeries, we may claim that there was a significant superiority when using the NeoSoniX handpiece in comparison to the conventional Legacy machine handpiece in cataract surgeries, regardless of cataract grade.
metadata.dc.description.unidade: Faculdade de Medicina (FMD)
Licença:: Arquivos Brasileiros de Oftalmologia - This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License (CC BY NC). Fonte: https://www.scielo.br/j/abo/a/bSL8LF9FSdvMS3bDX8qPV9B/?lang=pt#. Acesso em: 10 ago. 2021.
DOI: https://dx.doi.org/10.1590/S0004-27492006000300019
Aparece en las colecciones: Artigos publicados em periódicos e afins

Mostrar el registro Dublin Core completo del ítem " class="statisticsLink btn btn-primary" href="/jspui/handle/10482/26565/statistics">



Los ítems de DSpace están protegidos por copyright, con todos los derechos reservados, a menos que se indique lo contrario.