Título: Influência do macheamento do orifício piloto nos parafusos cervicais anteriores
Influence of tapping the pilot hole in anterior cervical screws
Influencia del taladramiento del agujero piloto en tornillos cervicales anteriores
Autor(es): Silva, Patrícia
Matheus, João Paulo Chieregato
Rosa, Rodrigo César
Porto, Maximiliano Aguiar
Paula, Francisco José Albuquerque de
Shimano, Antonio Carlos
Defino, Helton Luiz Aparecido
Assunto: Fraturas - fixação
Coluna vertebral
Parafusos ósseos
Histologia
Macheamento
Data de publicação: 2013
Editora: Sociedade Brasileira de Coluna
Referência: SILVA, Patrícia et al. Influência do macheamento do orifício piloto nos parafusos cervicais anteriores. Coluna/Columna, São Paulo, v. 12, n. 1, p. 21-24, 2013. Disponível em: <http://www.scielo.br/scielo.php?script=sci_arttext&pid=S1808-18512013000100004&lng=en&nrm=iso>. Acesso em: 21 maio 2018. doi: http://dx.doi.org/10.1590/S1808-18512013000100004.
Resumo: OBJETIVO: Analisar experimentalmente a influência do macheamento do orifício piloto nas propriedades mecânicas e na interface osso-implante do parafuso cervical anterior. MÉTODOS: Oito carneiros da raça Santa Inês deslanados foram utilizados no estudo. Nos segmentos vertebrais cervicais de C2-C7 foram realizados orifícios piloto de 2,5mm, de ambos os lados da vértebra. No lado direto da vértebra foi realizado o macheamento previamente a inserção do parafuso cortical de 3,5mm, e no lado esquerdo o implante foi colocado sem o macheamento do orifício piloto. Foram realizados ensaios mecânicos para avaliar a força de arrancamento dos implantes e estudo histomorfométrico da interface osso-implante na fase aguda e oito semanas após a colocação dos implantes. O torque de inserção foi mensurado durante a colocação dos implantes. RESULTADOS: O torque de inserção e a resistência ao arrancamento dos implantes foram maiores nos implantes colocados sem o macheamento do orifício piloto. A interface osso-implante apresentou maior área de contato ósseo e maior área óssea no interior da rosca do implante nos parafusos implantados sem o macheamento prévio. Não foi observado diferença na área óssea fora do passo de rosca do implante. CONCLUSÕES: O macheamento do orifício piloto reduz o torque de inserção e a resistência ao arrancamento do implante na fase aguda e crônica, e reduz a área de contato osso-implante e área de osso no interior do passo de rosca.
Abstract: OBJECTIVE: Experimentally analyze the influence of the pilot hole tapping on the mechanical properties and the bone-implant interface of anterior cervical screw. METHODS: Eight shorn Santa Inês sheep were used in the study. In cervical segments C2-C7 of the column 2.5mm pilot holes were made on both sides of the vertebra. On the right side of the vertebra tapping was done before the insertion of the cortical screw of 3.5mm, and on the left side the implant was placed without tapping the pilot hole. Mechanical assays were performed to assess the strength of pullout as well as a histomorphometric study of bone-implant interface during the acute phase and eight weeks after implant placement. The insertion torque was measured during implant placement. RESULTS: The insertion torque and pullout strength were higher for implants placed without tapping the pilot hole. The bone-implant interface showed greater area of bone contact and greater bone area within the implant thread on the screws implanted without prior tapping. No differences were observed in the bone outside the thread pitch of the implant. CONCLUSIONS: Tapping the pilot hole reduces insertion torque and pullout strength of the implant in both acute and chronic phases, and reduces the area of bone-implant contact and the bone area within the thread pitch.
Resumen: OBJETIVO: Para estudiar experimentalmente la influencia del taladramiento del agujero piloto sobre las propiedades mecánicas y la interfaz hueso-implante del tornillo cervical anterior. MÉTODOS: Ocho ovejas Santa Ines fueron utilizadas en el estudio. En los segmentos vertebrales cervicales C2-C7 se hicieron agujeros pilotos de 2,5 mm en ambos lados de la vértebra. En el lado derecho de la vértebra el taladramiento se hizo antes de la inserción del tornillo cortical de 3,5 mm, y en el lado izquierdo el implante fue colocado sin el taladramiento del agujero piloto. Se realizaron ensayos mecánicos para evaluar la fuerza de retirada y el estudio histomorfométrico de la interfase hueso-implante en la fase aguda y ocho semanas después de la colocación del implante. El torque de inserción se midió durante la colocación del implante. RESULTADOS: La fuerza de torsión de la inserción y la resistencia a la extracción fueron más altas en los implantes colocados sin el taladramiento del agujero piloto. La interfaz hueso-implante presentó mayor área de contacto con el hueso y mayor área de hueso dentro de la rosca del implante en los tornillos implantados sin taladramiento previo. No se observaron diferencias en el hueso fuera del paso de la rosca del implante. CONCLUSIONES: El taladramiento del agujero piloto reduce el par de torsión de inserción y la resistencia a la extracción del implante en las fases aguda y crónica, y reduce el área de contacto hueso-implante y el área ósea en el interior del paso de la rosca.
Licença: Coluna/Columna - All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License (CC By NC 4.0). Fonte: http://www.scielo.br/scielo.php?script=sci_arttext&pid=S1808-18512013000100004&lng=en&nrm=iso&tlng=pt. Acesso em: 21 maio 2018.
DOI: http://dx.doi.org/10.1590/S1808-18512013000100004